Spectroradiometer CS-2000/2000A

Spectroradiometer CS-2000/2000A
Features
  • Wide luminance measurement range.
  • Stable measurement ensured(PWM light sources).
  • Minimal polarization error.
  • Close-up lens for measurement of even tinier areas.
Specifications
CS-2000/CS-2000A
Wavelength range 380~780nm
Wavelength resolution 0.9nm/pixel
Display wavelength bandwidth 1.0nm
Wavelength precision ±0.3nm(Median wavelength : 435.8nm,546.1nm,643.8nm;Hg-Cd lamp)
Spectral bandwidth 5nm or less(half bandwidth)
Measuring angle(selectable) 0.2° 0.1°
Measurement luminance range(Standard light source A) CS-2000 0.003~5,000cd/m2 0.075~125,000cd/m2 0.3~500,000cd/m2
CS-2000A 0.0005~5,000cd/m2 0.0125~125,000cd/m2 0.05~500,000cd/m2
Minimum measuring area Ø5mm
(Ø1mm when using close-up lens)
Ø1mm
(Ø0.2mmmm when using close-up lens)
Ø0.5mm
(Ø0.1mmmm when using close-up lens)
Minimum measuring distance 350mm(55mm when using close-up lens)
Minimum measuring radiance display 1.0 x 10-9 W/sr x m2 x nm
Accuracy:Luminance(Standard light source A) ±2%
CS-2000 Accuracy:Luminance
(Standard light source A)
x,y:±0.0030 0.003~0.005cd/m2 0.075~0.125cd/m2 0.3~0.5cd/m2
x,y:±0.0020 0.005~0.05cd/m2 0.125~1.25cd/m2 0.5~5cd/m2
x:±0.0015 0.05cd/m2以上 1.25cd/m2以上 5cd/m2以上
y:±0.0010
Repeatability:Luminance(2σ)
(Standard light source A)
0.4% 0.003~0.05cd/m2 0.075~1.25cd/m2 0.3~5cd/m2
0.3% 0.05~0.1cd/m2 1.25~2.5cd/m2 5~10cd/m2
0.15% 0.1~5,000cd/m2 2.5~125,000cd/m2 10~500,000cd/m2
Repeatability:Chromaticity(2σ)
(Standard light source A)
x,y:0.0020 0.003~0.005cd/m2 0.075~0.125cd/m2 0.3~0.5cd/m2
x,y:0.0010 0.005~0.1cd/m2 0.125~2.5cd/m2 0.5~10cd/m2
x,y:0.0006 0.1~0.2cd/m2 2.5~5cd/m2 10~20cd/m2
x,y:0.0004 0.2~5,000cd/m2 5~125,000cd/m2 20~500,000cd/m2
CS-2000A Accuracy:Luminance
(Standard light source A)
x,y:±0.0015 0.001~0.05cd/m2 0.025~1.25cd/m2 0.1~5cd/m2
x:±0.0015 0.05cd/m2以上 1.25cd/m2以上 5cd/m2以上
y:±0.0010
Repeatability:Luminance(2σ)
(Standard light source A)
1.5% 0.0005~0.001cd/m2 0.0125~0.025cd/m2 0.05~0.1cd/m2
0.7% 0.001~0.003cd/m2 0.025~0.075cd/m2 0.1~0.3cd/m2
0.25% 0.003~0.05cd/m2 0.075~1.25cd/m2 0.3~5cd/m2
0.15% 0.05~5,000cd/m2 5~125,000cd/m2 5~500,000cd/m2
Repeatability:Chromaticity(2σ)
(Standard light source A)
x:0.0030
y:0.0035
0.001~0.003cd/m2 0.025~0.075cd/m2 0.1~0.3cd/m2
x:0.0030
y:0.0035
0.003~0.1cd/m2 0.075~2.5cd/m2 0.3~10cd/m2
x,y:0.0006 0.1~0.2cd/m2 2.5~5cd/m2 10~20cd/m2
x,y:0.0004 0.2~5,000cd/m2 5~125,000cd/m2 20~500,000cd/m2
Polarization error 1°:2% or less(400 to 780nm);0.1° and 0.2°:3% or less(400 to 780nm)
Integration time Fast : 0.005~16 sec/Normal : 0.005~120 sec.
Measurement time CS-2000 Approx. 2 sec. minimum(Manual mode) to 243 sec. maximum(Normal mode)
CS-2000A Approx. 2 sec. minimum(Manual mode) to 247 sec. maximum(Normal mode)
Color space Lv,x,y/Lv,u',v'/Lv,T,Δuv/X,Y,Z/spectral graph/dominant wavelength/excitation purity/scotopic luminosity(with CS-S10w Professional)
Interface USB1.1,RS-232C
Operation temperature/humidity range CS-2000 : 5 to 35°,relative humidity 80% or less with no condensation CS-2000A : 5 to 30°,relative humidity 80% or less with no condensation
Storage temperature/humidity range 0 to 35°,relative humidity 80% or less with no condensation
Power Dedicated AC Adapter (100-240 V~,50/60 Hz)
Current consumption Approx. 20W
Size(W x H x D) 156mm x 262mm x 392mm(Machine),Ø70mm x 95mm(Lens)
Weight Approx. 7.0Kg
Applications
Traffic lamp measurement
Car dashboard measurement
Outdoor Indicator lighting measurement
Display measurement